INTRODUCTION

Internal combustion engine emissions are comprised of a long list of organic compounds from C2 to C12 hydrocarbons. Major sources of these are automobiles and trucks with lesser sources including industrial emissions and home powered tools such as lawnmowers and leaf blowers. These emissions are considered as ozone precursors in the ambient air since under atmospheric conditions in the summer months, they can interact with nitrogen oxides in the presence of sunlight to produce ozone, a criteria air pollutant under the United States Clean Air Act (1970).

The amount and identity of the saturated and unsaturated organics that make up the ozone precursors has a major impact on the amount of ozone produced since each precursor has a certain ozone formation potential. Concentration at the tailpipe of an automobile are much greater than those found in the ambient air, so any analytical system must be capable of analysing a wide range of concentrations. A custom SCION GC system combining a special preconcentrator and dual FID was used to evaluate a standard gas mixture which contained target precursors from C2 to C13 hydrocarbons.

EXPERIMENTAL

A 456 GC was fitted with a sample preconcentration trap (SPT), a nafion gas sample dryer, a three Valco valve system with dual flame ionisation detectors (FID’s). Figure 1 shows the configuration of the gas analyser. Analytical parameters used throughout this application can be found in Table 1.

Approximately 100mL of sample is drawn through the dryer and then through the SPT multi-absorbent trap. The residual air is then flushed from the trap for one minute. The trap is then backflushed and desorbed at 220°C. The trap contents are split to two 0.32mm i.d capillary columns. Column 1, a PLOT Al2O3/KCl column separated C2 to C6 whereas a SCION-1 column separated C6 to C13. During the chromatography, the external dryer is backflushed with any trace of the sample being flushed out of the system.

RESULTS

A chromatogram is produced from each of the detectors; one for hydrocarbons C2 to C6 (Figure 2) and a second for hydrocarbons C6 to C13 (Figure 3).

A series of nine replicates of a C6 to C12 standard were run and the retention time and area count precisions determined. The statistical data for the precision testing can be found in Table 2.

DISCUSSION

Gas exhaust components or ozone precursors are targeted by EPA for quantitative analysis due to their ozone formation potential. However, there are a number of unidentified components that may be present in a sample. In order to correctly identify compounds, mass spectrometry may be used. Retention time, peak area t precision and the recovery of analytes is also important for the accurate quantification of the components of interest. Obtaining high recoveries of target compounds in air is more difficult as boiling point increases. Excellent recoveries are obtained through C11 (C6-C9 100%, C10 97%, C11 86%) with recovery dropping to 50% for C12

CONCLUSION

The SCION custom analyser with built in preconcentrator, dual columns and dual FID’s are capable of quantitatively analysing a wide range of hydrocarbon compounds found in vehicle emissions and ambient air. The high recovery and excellent precision ensure that targeted compounds are confidently identified and quantified.

Sriracha mi artisan farm-to-table bushwick health goth, fixie palo santo adaptogen. Affogato kale chips butcher shoreditch gastropub palo santo. Sartorial selvage sriracha polaroid forage umami. Sriracha locavore poke occupy, post-ironic quinoa live-edge chartreuse shabby chic ugh williamsburg cray everyday carry franzen.